Conventional protective measures against isolation faults, such as fault current circuit breaks for home installations, may not recognise potential isolation faults on the output side when isolation transformers are used. The ISOMONITOR monitors the dielectric resistances of both live contacts of the isolation transformer’s output socket against grounding potential and generates a warning signal in the event of a fault.
Isolation resistance is monitored in case the threshold value of 50 kΩ (25 kΩ) is exceeded. If this value is exceeded, both an acoustic signal (a pulsating tone of approx. 3 kHz, approx. 98 db) and a visual signal (LED indicator) are generated. The ISOMONITOR can be directly connected to one of the output sockets on the isolation transformer. Other optional functions are monitoring the transformer temperature with aural and visual alarms, and a visual display of the transformer’s power consumption.
In the event of a fault, the aural alarm can be reset with an acknowledgement button, and the visual alarm is automatically reset as soon as the fault has been dealt with. Should a function test be performed with the ISOMONITOR, an appropriate test plug (adapter cable required) for simulating an isolation fault must be plugged into one of the output sockets on the isolation transformer. The signal will be triggered while the test plug is plugged in.

ADVANTAGES

  • IEC 60601-1-2:2014
  • Easy to operate
  • Aural and visual signals
  • Function test with test plug
  • Power consumption display (optional)
  • Temperature warning (optional)
Mains voltage 230 115 (V)
Working range 200 – 240 100 -120 (VAC)
Threshold value ≤ 50 ≤ 25 (kΩ)
Response Time < 2 < 2 (sec.)
Signal Display In operation: LED green (Power on)
Insulation failure: LED yellow (Insulation)
Transformer limit temperature reached: LED yellow (temperature)
Transformer power load (Power): LED green: 30 % LED yellow: 60 % LED red: 90 %
Acoustic signal For insulation warning: pulsating continuous
At overtemperature: Continuous tone
Ambient operating temperature 0… +40 (°C)
relative ambient operating humidity 30… 75 (%)
Protection class II
Type of protection IP 20
Size [H x B x T] 192 x 34 x 56 (mm)

All data and configurations can be found in our product data sheet. Download datasheet